If you have any questions, please contact the following.
Check items | Check results (Date) | |||
---|---|---|---|---|
/ | / | / | / | |
1. More than 8 years have passed since original purchase of the illumination device, or the total using time exceeds 20,000 hours of use. | ||||
2. Lamp does not light sometimes even though the main switch is turned ON. (Only when using the halogen bulb) | ||||
3. Dirt or substances are attached around the main switch. | ||||
4. The light flickers if you move a lamp cable*1 or illumination devices. | ||||
5. Lamp cable*1 is unusually hot to the touch. | ||||
6. Burning or smoke odor. | ||||
7. The light still flickers even though the lamp is replaced
(Only when using the halogen bulb) | ||||
8. Signs of deformation, backlash, or looseness, etc. when you assemble/disassemble the illumination device.
(I.e. it is hard to open/ close the lid during lamp-replacement.) | ||||
9. Lamp connection terminals*2 have become discolored or tarnished, or the colors of the right and left terminals*3 are different.
(Only when using the halogen bulb) | ||||
10. Lamp cables*1 or wiring parts have become deformed, cracked or tarnished in any way. | ||||
11. Frequent repairs to similar devices put into use at the same time as the unit being checked. |
* If the spaces are not enough for check, copy this sheet.
*1 Discharge burners: Mercury burner / Xenon burner / Metal halide burner
*2 Example of a lamp cable. (Other models may not be shown here.)
Americas |
Patrick McGlone
Director, OSSA Technical Services Olympus Scientific Solutions Americas 3500 Corporate Parkway Center Valley, PA 18034 Tel: (1) 484-896-5255 E-mail: patrick.mcglone@olympus.com |
---|---|
Europe
Middle East Africa | Click here |
People's Republic of China |
OLYMPUS (China) Co.,LTD.Shanghai Branch
10F,K.W Center,1010 Huaihai Rd (M),Xuhui District, Shanghai,200031,China Scientific Solutions Business Division Quality & Regulatory Affairs Dept. Tel: (86)21-5170-6024 Fax:(86)21-5170-6236 E-mail: Changyi_Li@olympus.com.cn |
Japan | Customer support center |
Other Asia,Oceania | Please contact your nearest sales representative for Olympus microscopes. |
*3 Example of connection terminals. (Other models may not be shown here.)
*4 Example of a lamp installation terminal. (Other models may not be shown here.)
对不起,此内容在您的国家不适用。
您即将被转换到我们的本地网站。